Skc 225-380 Parallel Particle Impactor Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Zařízení Skc 225-380 Parallel Particle Impactor. SKC 225-380 Parallel Particle Impactor User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 8
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
1
SKC Re-usable Parallel Particle Impactors (PPI)
The patented
impaction-based SKC Re-usable Parallel Particle
Impactor (PPI ) Samplers are designed to match precisely the
collection e ciency curves for respirable and thoracic dust
speci ed by ISO/CEN 7708:1995 and adopted by ACGIH,
CEN, and other occupational hygiene organizations. The
performance of the respirable PPI samplers also meets
the speci cations in the 2013 OSHA Notice of Proposed
Rule Making (NPRM) for respirable crystalline silica. The
thoracic model meets the requirements of NIOSH Method
5524 for metalworking uids and compounds with ACGIH
®
thoracic TLV
®
s. Constructed of conductive aluminum, these
PPI Samplers may be re-used and o er a choice of ow rates
for maximum exibility in pump options, sample duration,
and contaminant concentration.
Performance Pro le
Sampling Rate: 2 L/min respirable or thoracic and 8 L/min
respirable
Sample Pump:Universal or AirChek
®
Series for 2 L/min
Leland Legacy for 8 L/min
Sample Time: Dependent on method used
Sample Media: 37-mm PVC lter, 5.0-μm pore size or
37-mm PTFE lter,* 2.0-μm pore size (NIOSH 5524) or
37-mm MCE lter, 0.8-μm pore size
Use cellulose pad or stainless steel screen for support.
Impaction
Substrate: Four 3/8-in diameter pre-oiled porous plastic discs, limited shelf-life
Analysis: Gravimetric or chemical
Body Material: Conductive aluminum
Dimensions:
Height: 4.25 in (10.8 cm) - clip to exhaust
Diameter: 1.8 in (4.6 cm)
Depth: 1.2 in (3.0 cm)
Weight: 3.3 oz (93.6 gm)
U.S. Patent No. 7,073,402
* Back pressure on PTFE lters can vary within the same lot.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]
Zobrazit stránku 0
1 2 3 4 5 6 7 8

Shrnutí obsahu

Strany 1 - Operating Instructions

1SKC Re-usable Parallel Particle Impactors (PPI)The patented† impaction-based SKC Re-usable Parallel Particle Impactor (PPI ) Samplers are designed to

Strany 2 - Media Preparation

2Principle of OperationSKC Re-usable PPI Samplers are impaction-based fi lter samplers that perform precise size-selection for either thoracic or resp

Strany 3 - Technical Tidbits:

34. Using forceps, insert a 37-mm support pad and a 37-mm collection fi lter into the base plate.Inserting Impaction Substrates into the PPI1. Ensure

Strany 4 - Calibration and Sampling

4Calibration and SamplingAs the particle load on the fi lter increases during sampling, the pressure drop will also increase. Therefore, use a compensa

Strany 5

53. Calibrate the sample pump to the specifi ed fl ow rate. See sample pump and calibrator operating instructions.4. Disconnect tubing and remove cali

Strany 6

6Removing the Collection Filter and Impaction Substrates1. Unscrew and remove the two screws that hold the inlet plate to the base plate.2. Li off

Strany 7 - References

7CleaningFor optimum performance, the PPI inlet, impaction, and base plates should be cleaned a er fi ve runs or upon a noticeable buildup of material

Strany 8 - Ordering Information

8Ordering InformationPPI Samplers, require fi lter, substrates, and support Cat. No.Respirable PPI (gold), 2 L/min 225-380Thoracic PPI (blue), 2 L/m

Komentáře k této Příručce

Žádné komentáře